KL-8081N
    Publish time 2022-04-01 21:35    
KL-8081N

Specifications


Pumping Mechanism

Curvilinear peristaltic

IV Set

Compatible with IV sets of any standard

Flow Rate

0.1-2000 ml/h

0.1099.99 mL/h  (in 0.01 ml/h increments)

100.0999.9 mL/h  (in 0.1 ml/h increments)

10002000 mL/h (in 1 ml/h increments)

Drops

1 drop/min -100drops/min  (in 1 drop/min increments)

Flow Rate Accuracy

±5%

Drop Rate Accuracy

±5%

VTBI

0.10mL99999.99mL (Minimum in 0.01 ml/h increments)

Volume Accuracy

<1 ml , ±0.2mL

>1ml, ±5 mL

Time

00:00:0199:59:59(h:m:s)  (Minimum in 1s increments)

Flow Rate  ( Body weight )

0.019999.99 ml/h (in 0.01 ml increments)

unit: ng/kg/minng/kg/hug/kg/minug/kg/hmg/kg/minmg/kg/hIU/kg/minIU/kg/hEU/kg/minEU/kg/h

Bolus Rate

Flow rate range : 502000 mL/h Increments

 (5099.99 )mL/h, (Minimum in 0.01mL/h increments)

100.0999.9mL/h, (Minimum in 0.1mL/h increments)

10002000mL/h, (Minimum in 1 mL/h increments)

Bolus Volume

0.1-50 ml (in 0.01 ml increments)

Accuracy: ±5% or ±0.2mL

Bolus , Purge

502000 mL/h (in 1 mL/h increments)

Accuracy: ±5%

Air Bubble Level

40800uL,adjustable.( in 20uL increments)

Accuracy: ±15uL or±20%

Occlusion Sensitivity

20kPa-130kPa, adjustable (in 10 kPa increments)

Accuracy: ±15 kPa or±15%

KVO Rate

1).Automatic KVO on/ off function

2).Automatic KVO is turned off : KVO Rate : 0.110.0 mL/h adjustable(Minimum in 0.1mL/h increments).

When flow rate>KVO rate , it runs in KVO rate.

When flow rate

3) Automatic KVO is turned on :it adjust the flow rate automatically.

When flow rate<10mL/h, KVO rate =1mL/h

When flow rate >10 mL/h, KVO=3 mL/h.

Accuracy: ±5%

Basic function

Dynamic Pressure Monitoring, Key Locker, Standby, Historical memory, Drug Library.

Alarms

Occlusion, air-in-line, door open, near end , end program, low battery, end battery, motor malfunction, system malfunction, drop error, standby alarm

Infusion Mode

Rate mode, Time mode, Body weight, Sequence ModeDose ModeRamp Up/Down ModeMicro-Infu ModeDrop Mode.

Additional Features

Self-checking, System Memory, Wireless (optional),Cascade, Battery Missing Prompt, AC Power Off Prompt.

Air-in-line Detection

Ultrasonic detector

Power Supply, AC

AC100V240V 50/60Hz, 35 VA

Battery

14.4 V, 2200mAh ,Lithium, rechargeable

Weight of Battery

210g

Battery Life

10 hours at 25 ml/h

Working Temperature

5~40

Relative Humidity

15%80%

Atmospheric Pressure

86KPa106KPa

Size

240×87×176mm

Weight

<2.5 kg

Safety Classification

Class ⅠI, type CF. IPX3